Getinge Usa Sales Inc recalls Vaporizer Sevoflurane, Quik-Fil. Model Number: 6682285. Dev

Recalled Product

Vaporizer Sevoflurane, Quik-Fil. Model Number: 6682285. Device is a component of Flow-i C20, Flow-i C30, Flow-i C40, Flow-c, Flow-e. For use during anesthesia.


Hazard / Issue

Sevoflurane used with the vaporizer may degrade to hydrogen fluoride, which could present a risk to patients or health care providers through inhalation and/or skin exposure. Potential harms include irritation of respiratory tract, which in worst case may lead to lung edema and/or severe hypocalcemia, blistering, superficial ulceration, and/or hypomagnesemia.

Issued by

FDA

Distribution: Domestic distribution to Kentucky and Connecticut.
Lot/Code Info: Model Number: 6682285. UDI-DI: 07325710001141. Serial Numbers: 3761, 7635, 7633, 7630, 7632, 7631.