medical Class II Updated 2024-02-21

Microbiologics Inc recalls Microbiologics Enteric Pathogens Control Panel (Inactivated

Recalled Product

Microbiologics Enteric Pathogens Control Panel (Inactivated Pellet), REF 8184.


Hazard / Issue

The negative control was contaminated with one of the positive control pathogens (Norovirus). This would result in users getting a positive call for Norovirus when running the negative control.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Lot #8184-68, Exp. 5/31/2025, UDI-DI 70845357041448. UPDATE: The recall was expanded on 1/23/2024 to include the following additional lot numbers: 8184-65, exp. 3/31/2025; 8184-66, exp. 4/30/2025; 8184-67, exp. 5/31/2025; 8184-69, exp. 6/30/2025; 8184-70, exp. 8/31/2025; and 8184-71, exp. 8/31/2025.
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →