medical Class II Updated 2025-02-12

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD10 Catalog (1) 722003 (2) 722010 (3)722

Recalled Product

Allura Xper FD10 Catalog (1) 722003 (2) 722010 (3)722026


Hazard / Issue

Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: UDI: (1) 722003 00884838059191(10); (2) 722010 00884838059030(21); (3)722026 00884838054189(21).
View official government recall