medical Class II Updated 2022-05-25

Beckman Coulter Inc. recalls MicroScan WalkAway-96 plus Instrument REF B1018-284

Recalled Product

MicroScan WalkAway-96 plus Instrument REF B1018-284


Hazard / Issue

WalkAway Instruments may have been assembled with the incorrect AC power entry connector inlet with is rated at 1 ampere current rating rather than the correct current rating of 10 ampere

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Catalog Number: B1018-284 UDI-DI Code: 15099590658687 Serial Numbers: 39610007 39610008 39610009 39610010 39610011 39610012 39610013
View official government recall