medical Class II Updated 2022-05-25

Beckman Coulter Inc. recalls MicroScan WalkAway-40 plus Reconditioned Instrument REF B101

Recalled Product

MicroScan WalkAway-40 plus Reconditioned Instrument REF B1018-340R


Hazard / Issue

WalkAway Instruments may have been assembled with the incorrect AC power entry connector inlet with is rated at 1 ampere current rating rather than the correct current rating of 10 ampere

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Catalog Number: B1018-340R UDI-DI Code: 15099590684112 Serial Numbers: 34020035 34020224 34030054