medical Class II Updated 2022-05-25

Beckman Coulter Inc. recalls MicroScan WalkAway-96 plus Reconditioned Instrument REF B101

Recalled Product

MicroScan WalkAway-96 plus Reconditioned Instrument REF B1018-396R


Hazard / Issue

WalkAway Instruments may have been assembled with the incorrect AC power entry connector inlet with is rated at 1 ampere current rating rather than the correct current rating of 10 ampere

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Catalog Number: B1018-396R UDI-DI Code: 15099590684129 Serial Numbers: 39620362 39620406 3967731 3967914
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →