medical Class II Updated 2025-02-12

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Azurion 7 B12 System Code: (1) 722067 (2) 722225 (3)

Recalled Product

Azurion 7 B12 System Code: (1) 722067 (2) 722225 (3) 722235


Hazard / Issue

Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: UDI: (1) 722067 00884838085350(21); (2) 722225 00884838099265(21); (3) 722235 00884838116788(21)
View official government recall