medical Class III Updated 2022-05-25

Randox Laboratories Ltd. recalls CK MB Calibrator, Catalogue Number CK2393

Recalled Product

CK MB Calibrator, Catalogue Number CK2393


Hazard / Issue

Randox have realigned CK-MB Calibrator, CK2393, lot number 4410CK. Calibrating with the new value will cause a positive shift of approximately 6% with both Quality Controls and Patient Samples.

Issued by

FDA

Affected States: WV
Lot/Code Info: GTIN 05055273201451 Lot 4410CK