medical Class II Updated 2021-02-17

Draegar Medical Systems, Inc. recalls Infinity Acute Care System (IACS) Monitoring Solution: The

Recalled Product

Infinity Acute Care System (IACS) Monitoring Solution: The IACS is intended for multi-parameter, physiologic patient monitoring of adult, pediatric and neonatal patients in environments where patient care is provided by trained healthcare professionals.


Hazard / Issue

The Infinity M540 patient monitor may randomly reboot due to an error to correctly transmit and read the header data of files in the memory of the device. Under this situation, the device will try to reboot to mitigate the error. The device will be available again for use within 30 seconds. If this error continues and the M540 reboots three times in a time span of 10 minutes, it will enter a fail-state. A fail-state will annunciate itself with a continuous sound to alert the user. The M540 will reset to factory default and the user can manually configure and readmit the patient to continue patient monitoring.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Model Number - MS25510, MS25520, MS26372 Revisions 24 to 29
View official government recall