medical Class II Updated 2020-02-19

bioMerieux, Inc. recalls VIDAS Varicella Zoster IgG (VZG) antibody IVD, kit

Recalled Product

VIDAS Varicella Zoster IgG (VZG) antibody IVD, kit


Hazard / Issue

Two batches of VIDAS Varicella-Zoster IgG (VZG ref 30217) may yield an invalid calibration which may delay the patient's sample result or a valid calibration may lead to a false positive patient result.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: US Model 30217-01 Lot 1007410330 and 1007393380
View official government recall