medical Class II Updated 2022-05-25

Luminex Corporation recalls VERIGENE Nucleic Acid Test Kit-BC-GN Extraction Tray - 9.73

Recalled Product

VERIGENE Nucleic Acid Test Kit-BC-GN Extraction Tray - 9.73 mL, REF 20-009-021


Hazard / Issue

Higher than normal "No Call Internal Control failures" due to a missing reagent within a small subset of the extraction tray lot.

Issued by

FDA

Distribution: US
Lot/Code Info: Lot Number 101321021B