Luminex Corporation recalls VERIGENE Nucleic Acid Test Kit-BC-GN Extraction Tray - 9.73
Recalled Product
VERIGENE Nucleic Acid Test Kit-BC-GN Extraction Tray - 9.73 mL, REF 20-009-021
Hazard / Issue
Higher than normal "No Call Internal Control failures" due to a missing reagent within a small subset of the extraction tray lot.
Issued by
FDA
Distribution: US
Lot/Code Info: Lot Number 101321021B