medical Class II Updated 2019-04-17

PROCEPT BIOROBOTICS CORPORATION recalls AquaBeam Handpiece, Sterile EO, Rx Only, REF: 310301, and 32

Recalled Product

AquaBeam Handpiece, Sterile EO, Rx Only, REF: 310301, and 320301, a component of AquaBeam Robotic System. 320301 distributed outside the U.S. Device Description The AQUABEAM Robotic System is intended for use in patients suffering from lower urinary tract symptoms resulting from benign prostatic hyperplasia (BPH).


Hazard / Issue

Certain lots may develop an internal leak due to a micro-crack in a specific weld joint within the device. When the Robot detects this failure, it immediately renders the device inoperable and notifies the user via an error message, which may cause delay and/or inability to complete the procedure.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Domestic Lots: 18C00681, 18C00770, 18C00852, 18C01244. Foreign Lots: 18C00799, 18C00844, 18C00985, 18C01024, 18C01044, 18C01234