medical Class I Updated 2019-05-01

Edwards Lifesciences, LLC recalls Edwards Lifesciences, Miller Balloon Atrioseptostomy Cathete

Recalled Product

Edwards Lifesciences, Miller Balloon Atrioseptostomy Catheter, Sterile EO, Rx Only, REF: 830515F Product Usage: The Miller balloon atrioseptostomy catheter is designed for enlarging interatrial openings. In most pediatric cardiology centers, balloon atrioseptostomy is an accepted technique for palliation of several congenital cardiac defects. Description of Use Balloon atrioseptostomy can be performed in conjunction with diagnostic cardiac catheterization or under echocardiographic guidance and has been carried out after the diagnosis of several congenital cardiac defects including but not limited to: transposition of the great arteries, total anomalous pulmonary venous drainage without pulmonary obstruction, tricuspid atresia, and pulmonary atresia with intact ventricular septum.


Hazard / Issue

Difficulty with atrioseptostomy catheter balloon deflation after deployment, and reports of balloon fragmentation or detachment.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: All lots within 15 months expiry from 03/01/2019