medical Class II Updated 2022-06-01

International Science & Technology, LP, DBA Diamatrix Ltd. recalls Outer Packaging Label: Protekt Sharps Safety Knives - REF:

Recalled Product

Outer Packaging Label: Protekt Sharps Safety Knives - REF: PG 44620MU, Sterile R, Multi-Use 2.2mm Double Bevel Slit, Angled 6 units, Individual Packaging Label: ProTekt Sharps Safety Knives - REF: PG 44620MU,2.2mm double Bevel Slit, Sterile R, Angled 1 unit, Handle Labeling; 2.5mm slit double bevel


Hazard / Issue

Surgical knives may contain incorrect description printed on handles. Specifically, some surgical knife handles may contain a size of "2.5mm" when the actual size is "2.2mm".

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Item # PG 44620MU; Lot # : B923LUC