medical Class II Updated 2021-03-03

Ortho-Clinical Diagnostics, Inc. recalls VITROS Chemistry Products VersaTips - Product Usage: used to

Recalled Product

VITROS Chemistry Products VersaTips - Product Usage: used to dispense samples to MicroSlide assays and sample and reagent fluid for MicroWell assays. In addition, VersaTips are used to dispense R1 Reagent Fluid into cuvettes in combination with sample fluid for Micro Tips Assays.


Hazard / Issue

A deformity affecting a specific lot of VITROS¿ Chemistry Products VersaTips. Deformed VersaTips results may be biased by up to +/-16% for both immunometric and competitive assays.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Lot 413420 Model(s): VITROS 250, VITROS 350, VITROS ECi, VITROS 5,1 FS, VITROS 3600, VITROS 4600, VITROS 5600, VITROS XT7600.
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →