medical Class II Updated 2021-03-03

Natus Manufacturing Limited recalls Ref: NT821731C, Natus EDS 3 CSF External Drainage System - P

Recalled Product

Ref: NT821731C, Natus EDS 3 CSF External Drainage System - Product Usage: indicated for draining cerebrospinal fluid (CFS) from the cerebral ventricles or the lumbar subarachnoid space as a means of reducing intracranial pressure and CSF volume when the insertion of a permanent, internal shunt is not indicated.


Hazard / Issue

The external drainage system failed to meet Pyrogen test requirements for devices contacting cerebrospinal fluid of 2.15 endotoxins.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Lot # 118000422298
View official government recall