medical Class II Updated 2020-02-19

Pacific Medical Group Inc. recalls Toco+ transducer with ECG/IUP capability (wired), that may h

Recalled Product

Toco+ transducer with ECG/IUP capability (wired), that may have been serviced with the following: Pacific Medical Philips Avalon Smart US/TOCO Cable Assembly, NFPHA9020; Pacific Medical Philips Avalon Smart M2734A/2735A Toco Gauge Flex, NFPHA9341


Hazard / Issue

Fetal/maternal monitor/monitoring systems serviced or remanufactured using non-OEM equivalent components that have not been appropriately verified or validated. Possible adverse clinical impact: inability to accurately detect and measure fetal heart rate, inability to accurately detect maternal peripheral oxygen saturation, inability to monitor uterine activity to follow contractions, inability to determine temperature, and maternal blood pressure, lack of electrocardiography tracing, patient/clinician burns, and delay in detecting maternal or fetal distress.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: All serial numbers serviced between 9/2/15 and 4/2/19.