medical Class II Updated 2024-02-28

FUJIFILM Healthcare Americas Corporation recalls Synapse Cardiology PACS V7.3.0, V7.2, V7.1, V7.0. A web-ba

Recalled Product

Synapse Cardiology PACS V7.3.0, V7.2, V7.1, V7.0. A web-based application as the primary user interface for the processing of medical images.


Hazard / Issue

The LV Mass (2D Bullet) equation may be calculated incorrectly, resulting in a variance in the mass of the left ventricle (LV Mass). If used as a main factor for diagnosis, there is a risk of misdiagnosis or incorrect treatment plan of a patient, resulting in long-term health consequences or serious deterioration of health.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: UDI-DI 854904006008 Software versions 7.3.000, 7.2.200, 7.2.100, 7.2.000, 7.1.000, 7.0.000
View official government recall