medical Class II Updated 2025-02-26

RAYSEARCH LABORATORIES AB recalls Brand Name: RayStation- Intended as a Radiation Therapy Trea

Recalled Product

Brand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 10.1.0.613 and 10.1.1.54 Software Version: RayStation 10B and 10B Service Pack 1


Hazard / Issue

Inconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT calibration curve used in RayStation version 4.5 to 2024B including some service packs.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Lot Code: UDI: 0735000201031020201216 and 0735000201047120220128 GTIN: 07350002010310 and 07350002010471 Serial Numbers: 10.1.0.613 and 10.1.1.54 Software Revisions: RayStation 10B and 10B Service Pack 1 Expiration Date: 2027-02-01