medical Class II Updated 2025-02-26

RAYSEARCH LABORATORIES AB recalls Brand Name: RayStation- Intended as a Radiation Therapy Trea

Recalled Product

Brand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 12.0.0.932, 12.1.0.1221, 12.0.3.68, 12.0.4.12 and 12.3.0.119 Software Version: RayStation 11B and 11B Service Pack 1, 2, 3 and Service Pack Toshiba 1


Hazard / Issue

Inconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT calibration curve used in RayStation version 4.5 to 2024B including some service packs.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Lot Code: UDI: 0735000201042620211208, 0735000201049520220312, 0735000201050120220422, 0735000201060020220620 and 0735000201057020221222 GTIN: 07350002010426, 07350002010495, 07350002010501, 07350002010600 and 07350002010570 Serial Numbers: 12.0.0.932, 12.1.0.1221, 12.0.3.68, 12.0.4.12 and 12.3.0.119 Software Revisions: RayStation 11B and 11B Service Pack 1, 2, 3 and Service Pack Toshiba 1 Expiration Date: 2028-03-20