GE Healthcare, LLC recalls Innova IGS 630, Angiographic X-Ray, s5i system option used
Recalled Product
Innova IGS 630, Angiographic X-Ray, s5i system option used in generating fluoroscopic and rotational images of human anatomy for cardiovascular, vascular and non-vascular, diagnostic and interventional procedures.
Hazard / Issue
There is a potential for loss of the x-ray imaging function when the user changes field of view (FOV) from 30cm to 20cm or from 20cm to 30cm while releasing the fluoroscopy footswitch pedal simultaneously.