medical MODERATE Updated 2025-02-26

Dexcom, Inc. recalls Dexcom G6, Continuous Glucose Monitoring System, REF: STK-OE

See all recalls from Dexcom, Inc. →

Recalled Product

Dexcom G6, Continuous Glucose Monitoring System, REF: STK-OE-001

Share this recall: Facebook X / Twitter WhatsApp Email

Hazard / Issue

Under very rare situations, the Dexcom G6 touchscreen receiver may not provide high or low glucose alarms/alerts as designed, which can result in two different conditions. First, if a receiver operating system .net error occurs when an alarm/alert should be triggered, the initial alarm/alert will not be delivered until a subsequent alarm/alert is triggered. At that time, the initial alarm/alert is delivered, but not the second one. Alarm/alerts will continue to be delayed and be one alarm/alert behind, causing delayed alarm/alerts until the receiver is reset. This condition could result in the missed detection of a hyperglycemic or hypoglycemic event. Second, a single EGV reading may be delayed for 5 minutes after the initial .net operating system error. If the user receives a single EGV reading delayed by 5 minutes, the delay is not expected to cause user harm.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: All Serial Numbers/UDI: 00386270001900

✅ What Should You Do?

  1. 1Contact your healthcare provider immediately to discuss this recall.
  2. 2Stop using the device if it is safe to do so, and contact your provider.
  3. 3Follow the manufacturer's remedy instructions provided in the recall notice.
  4. 4Report device problems to FDA MedWatch or call 1-800-FDA-1088.

Looking for a safe alternative?

Search Amazon for Dexcom G6, Continuous Glucose →
View official government recall