medical Class I Updated 2022-06-15

GE Healthcare, LLC recalls Replacement back up batteries distributed on or after April

Recalled Product

Replacement back up batteries distributed on or after April 1, 2019, for CARESCAPE R860, Engstr¿om Carestation and Engstr¿om PRO Ventilators.


Hazard / Issue

Insufficient battery backup power resulting in premature shutdown of the ventilator when not connected to AC mains power supply.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Field Replacement Unit kits PN: 1009-5682-000-S (BTRY SEALED LEAD ACID RECHARGEABLE 12V) and PN: 5856787-S (BTRY SEALED LEAD ACID RECHARGEABLE 12V PAIR).
View official government recall