medical Class II Updated 2020-02-26

Siemens Healthcare Diagnostics, Inc. recalls Atellica CH Enzymatic Creatinine (ECRE_2) reagent, SMN 11097

Recalled Product

Atellica CH Enzymatic Creatinine (ECRE_2) reagent, SMN 11097533


Hazard / Issue

There is a potential for falsely depressed creatinine results for patients on phenindione therapy when using the enzymatic methodology. Interference has not been observed with the creatinine Jaffe methodology.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: UDI Number: 00630414595795 Lots: All lots
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →