medical Class II Updated 2020-02-26

Siemens Healthcare Diagnostics, Inc. recalls Dimension Enzymatic Creatinine (EZCR) reagent, SMN 10471520

Recalled Product

Dimension Enzymatic Creatinine (EZCR) reagent, SMN 10471520


Hazard / Issue

There is a potential for falsely depressed creatinine results for patients on phenindione therapy when using the enzymatic methodology. Interference has not been observed with the creatinine Jaffe methodology.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: UDI Number: 00842768030512 Lots: All lots