medical Class II Updated 2020-02-26

Siemens Healthcare Diagnostics, Inc. recalls Dimension Vista Enzymatic Creatinine (ECREA) reagent, SMN 10

Recalled Product

Dimension Vista Enzymatic Creatinine (ECREA) reagent, SMN 10700444


Hazard / Issue

There is a potential for falsely depressed creatinine results for patients on phenindione therapy when using the enzymatic methodology. Interference has not been observed with the creatinine Jaffe methodology.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: UDI Number: 00842768036262 Lots: All lots
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →