medical Class II Updated 2019-05-08

Philips Medical Systems Nederlands recalls Ingenia Ambition S (781359), Ingenia Ambition X (781356) wit

Recalled Product

Ingenia Ambition S (781359), Ingenia Ambition X (781356) with Magnet Energization Device (MED), Philips MR systems are indicated for use as a diagnostic device. The purpose of the MED is to supply power to charge the magnet. MED is abbreviation for Magnet Energize Device


Hazard / Issue

Magnet Energization Device (MED) can have a residual voltage on output terminal, may cause electrical shocks to personnel servicing the MED, MDD or connected cabling.

Issued by

FDA

Affected States: FL
Lot/Code Info: US Serial Number: 47004 OUS Serial Numbers: 47001, 47002, 48000
View official government recall