medical Class II Updated 2026-02-11

Focalyx Technologies, LLC. recalls Focalyx Fusion

Recalled Product

Focalyx Fusion


Hazard / Issue

Software device that is an accessory for image-guided interventional and diagnostic procedures involving prostate has accessories that may not function properly with Windows 10, which may cause patient harm during biopsy or cause treatment/therapy delays, so firm is recommending stop use until device is verified/validated with Windows 11.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: UDI-DI: 00860003552001. Software Versions: 1.3, 1.5, 1.6. Serial Numbers: IF-FF-B-1-22080504, IF-FF-B-1-TPAA998118, IF-FF-B-1-351238, IF-FF-B-1-TPAB081335, IF-FF-B-1-22080508, IF-FF-B-1-TPAB200331, IF-FF-B-1-TPAB200333, IF-FF-B-1-TPAC505383, IF-FF-B-1-TPAB200348, IF-FF-B-1-TPAB540364, IF-FF-B-1-S5ARQL009253NYZ, IF-FF-B-1-SCARQL0014096YZ, IF-FF-B-1-TPAB532163, IF-FF-B-1-SCARQL00136033KK, IF-FF-B-1-TPAB540359, IF-FF-B-1-TPAB081326, IF-FF-B-1-22080505, IF-FF-B-1-TPAB081324