Cardinal Health 200, LLC recalls Magellan Hypodermic Safety Needle, 25 G x 5/8" (0.508 mm x 1
Recalled Product
Magellan Hypodermic Safety Needle, 25 G x 5/8" (0.508 mm x 1.6 cm). A single lumen needle intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. Product Code: 8881850558
Hazard / Issue
Safety shield not locking into the cannula. If the safety shield activation fails to lock a contaminated needle, a needlestick may occur and there is the potential for exposure to bloodborne pathogens
Issued by
FDA