medical MODERATE Updated 2018-04-04

Encore Medical, Lp recalls Altivate Anatomic(TM) INSTRUMENTATION, PEGGED GLENOID TRIAL,

See all recalls from Encore Medical, Lp →

Recalled Product

Altivate Anatomic(TM) INSTRUMENTATION, PEGGED GLENOID TRIAL, SIZE 42mm, REF 804-07-420 Product Usage: The AltiVate Anatomic Shoulder System is a total shoulder system indicated as an anatomic shoulder joint replacement for patients suffering from pain and dysfunction.

Share this recall: Facebook X / Twitter WhatsApp Email

Hazard / Issue

It was discovered during a surgery that the design of the radius at the base of the pegs of the Anatomic Glenoid Trial (Part Numbers: 804-07-380, 804-07-420, 804-07-460, 804-07-500, and 804-07-540) is much larger on the trails than on the implants

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: All Serial Numbers