medical Class II Updated 2020-02-26

Siemens Healthcare Diagnostics, Inc. recalls ADVIA Chemistry CardioPhase High Sensitivity C-Reactive Prot

Recalled Product

ADVIA Chemistry CardioPhase High Sensitivity C-Reactive Protein (hsCRP) Calibrator - product Usage: For in vitro diagnostic use in the calibration of ADVIA¿ Chemistry systems for the CardioPhase¿ hsCRP assay.


Hazard / Issue

Atellica and ADVIA hsCRP calibrator accuracy has drifted away from the reference materials (ERM-DA470/IFCC), which is causing a high bias on QC and patient sample recovery.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Siemens Material Number (SMN): ADVIA Chemistry: 10335897 REF: 05006455 Lot# 453610 Expiration Date: 2020-01 UDI Number: ADVIA: (01)00630414526843(10)453610(17)20200128
View official government recall