medical Class II Updated 2022-07-06

Microgenics Corporation recalls QMS Tacrolimus Immunoassay REF 10015573 CALIBRATOR SET QM

Recalled Product

QMS Tacrolimus Immunoassay REF 10015573 CALIBRATOR SET QMS TACROLIMUS CAL A 10015566-3 QMS TACROLIMUS CAL B 10015567-3 QMS TACROLIMUS CAL C 10015568-3 QMS TACROLIMUS CAL D 10015569-3 QMS TACROLIMUS CAL E 10015570-3 QMS TACROLIMUS CAL F 10015571-3


Hazard / Issue

Due to QMS Tacrolimus Calibrator recovering between 82 to 94% of their target values, resulting in over-quantitation in tacrolimus patient result values when used with the QMS Tacrolimus Immunoassay.

Issued by

FDA

Affected States: AL, AZ, CA, DC, FL, IA, ID, IL, KY, LA, MA, NE, NJ, NM, OH, OK, PA, SC, TN, TX, VA
Lot/Code Info: Model/Part Number: 10015573 UDI Codes: (01) 0884883 01259 2 (17) 220310 (10) 74175462 (01) 0884883 01259 2 (17) 220512 (10) 74267412 Lot Numbers: 74175462 74267412 Calibrators within the set: QMS TACROLIMUS CAL A 10015566-3 QMS TACROLIMUS CAL B 10015567-3 QMS TACROLIMUS CAL C 10015568-3 QMS TACROLIMUS CAL D 10015569-3 QMS TACROLIMUS CAL E 10015570-3 QMS TACROLIMUS CAL F 10015571-3