medical Class II Updated 2019-05-15

Medica Corporation recalls Medica Wash1-Wash solution used prior to Lipase assay on the

Recalled Product

Medica Wash1-Wash solution used prior to Lipase assay on the EasyRA clinical chemistry analyzer REF 10680 - Product Usage: For use in washing the probe of the Medica EasyRA Chemistry Analyzer. The wash solution in the wedge is used to prevent interference between certain tests.


Hazard / Issue

Wash1 solution is ineffective in cleaning dispensing probe of TRIG and LDL reagents that interfere with the Lipase (LIP) assay on the EasyRA clinical chemistry analyzer.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Lot #s 18066, 18133, 18235, 18317