Siemens Medical Solutions USA, Inc recalls Cios Alpha, Model Number 10308191 Mobile X-ray system - Pro
Recalled Product
Cios Alpha, Model Number 10308191 Mobile X-ray system - Product Usage: The Cios Alpha is a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications. Clinical applications may include but are not limited to interventional fluoroscopic, gastro-intestinal, endoscopic, urologic, pain management, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures. The patient population may include pediatric patients.
Hazard / Issue
Continuous operation at very high tube output may result in increased wear of the radiation emitting components. This may occur when tube voltages of 125 kV are used for a long period, e.g. 30 minutes or more of continuous use. As a result, risk of premature failure of the Monoblock and loss of imaging X-rays.