medical Class II Updated 2022-07-06

Baxter Healthcare Corporation recalls NaviCare Nurse Call/Voalte Nurse Call

Recalled Product

NaviCare Nurse Call/Voalte Nurse Call


Hazard / Issue

An issue has been identified with Phillips (Emergin) and Longleaf non relay (Connexall, Vocera, Cerner) wireless integrations used with NaviCare/Voalte Nurse Call, software versions 3.9.100 through 3.9.300. Calls placed from a push button call device, such as a push button switch, call cord, or from the push buttons on a room audio station, will be canceled on the nurse call system when the call is answered at the wireless phone, regardless of the call priority.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: software versions: 3.9.100, 3.9.101,3.9.102, 3.9.200, 3.9.201, and 3.9.300
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →