medical Class I Updated 2021-04-07

CareFusion 303, Inc. recalls BD Alaris Infusion Pump Module Model 8100 Pump Module Keypad

Recalled Product

BD Alaris Infusion Pump Module Model 8100 Pump Module Keypad; and Pump Module Door Assembly Replacement Kit with Pump Module Keypad, Parts: 49000239; 49000346; 49000438; 49000439.


Hazard / Issue

Pump Module keypad lifting, and Fluid ingress could result in 1) Unresponsive keys: module continues infusion, PC unit will not alarm and must be used for programming changes, may necessitate a different Pump Module. May result in infusion start delay/inability to titrate medication. 2) Stuck keys: PC unit alarms, module exhibits Channel Error, may result in infusion interruption/start delay

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Pump Module with Keypad: manufactured January 23, 2019 to December 5, 2019. Pump Module Door Assembly Kit with Keypad: Dated from January 15, 2019 to November 14, 2019.
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →