medical Class II Updated 2026-02-18

Philips North America Llc recalls IQon Spectral CT; Product Code (REF): 728332;

Recalled Product

IQon Spectral CT; Product Code (REF): 728332;


Hazard / Issue

As part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry. One or more of these fasteners may not be torqued to specification. If multiple fasteners are not torqued to specification, the rotor or rotor parts may become unsecured or displaced. While there has been no reported or observed event of expelled parts during gantry rotation, unsecured rotor parts may potentially be expelled.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Product Code (REF): 728332; UDI-DI: 00884838059542; Serial Numbers: 60128, 60141, 60151, 60161, 60171, 60181, 60129, 60142, 60152, 60162, 60172, 60182, 60130, 60143, 60153, 60163, 60173, 60132, 60144, 60154, 60164, 60174, 60133, 60145, 60155, 60165, 60175, 60134, 60146, 60156, 60166, 60176, 60136, 60147, 60157, 60167, 60177, 60137, 60148, 60158, 60168, 60178, 60138, 60149, 60159, 60169, 60179, 60139, 60150, 60160, 60170, 60180,