medical Class I Updated 2026-02-18

Medline Industries, LP recalls Medline ReNewal Reprocessed Webster CS Catheters: Webster

Recalled Product

Medline ReNewal Reprocessed Webster CS Catheters: Webster CS Catheter 10 Electrodes 2-8-2mm Spacing w/AutoID w/ EZ Steer BD710DF282CRH; Webster CS Catheter 10 Electrodes, 2-8-2mm Spacing w/ EZ Steer BD710DF282RRH BD710FJ282RRH; Webster CS Catheter 10 Electrodes, 2-8-2mm Spacing w/AutoID w/ EZ Steer BD710FJ282CRH;


Hazard / Issue

These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus reaction which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Expansion of recall Z-2610/2614-2025

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: BD710DF282CRH UDI-DI 10197344044023 Lots EP250429 EP250808; BD710DF282RRH UDI-DI 10197344044030 Lots EP250429 EP250808; BD710FJ282RRH UDI-DI 10197344044054 Lots EP250429 EP250808; BD710FJ282CRH UDI-DI (case) 40197344044048 (ea) 10197344044047 Lots EP250429 EP250808
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →