medical Class II Updated 2019-05-22

Philips North America, LLC recalls SureSigns VS3 NBP, SpO2, Temp, Product Number 863073. For m

Recalled Product

SureSigns VS3 NBP, SpO2, Temp, Product Number 863073. For monitoring the physiological parameters of patients (NBP, SpO2 & Temperature).


Hazard / Issue

System software inhibits the monitor and as a result does not measure, display and alarm for pulse rates above 240 beats per minute.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Serial Numbers: All Including UDIs: (01)00884838028890(21) and (01)00884838087075(21) followed by serial number.
View official government recall