PENTAX of America Inc recalls Pentax Medical C2 CryoBalloon Pear Focal Catheter, Model FG

Recalled Product

Pentax Medical C2 CryoBalloon Pear Focal Catheter, Model FG 1024, Sterile, RX. Product Usage: The C2 CryoBalloon Ablation System is intended to be used as a cryosurgical tool in the field of general surgery, specifically for endoscopic application and the ablation of dysplastic Barrett s Esophagus. The system includes the following components: (1) CryoBalloon Catheter connects to the Controller, which controls the operation of the Catheter such as diffuser (sprayer) positioning and ablation (nitrous oxide release); (2) CryoBalloon Controller contains the cartridge heater and cryogen delivery valve, which is controlled with the foot pedal; (3) Foot Pedal; and (4) CryoBalloon Cartridge containing 36 grams of nitrous oxide.


Hazard / Issue

Incorrect default dose and dose increments may be transmitted to the controller if the catheter RFID tag is not correct or is not being correctly or completely readable by the controller.

Issued by

FDA

Affected States: CA, DC, FL, LA, MA, MD, MN, NC, NJ, NY, OH, PA, SC, TN, TX, VA, WA
Lot/Code Info: Lot/Serial numbers: 06152018-01, 07172018-02, 07022018-04, 08212018-01, and 08102018-03.