medical Class II Updated 2022-07-13

Siemens Medical Solutions USA, Inc recalls Artis zeego (Model no. 10280959) and Artis Q.zeego (Model n

Recalled Product

Artis zeego (Model no. 10280959) and Artis Q.zeego (Model no. 10848283), Interventional Fluoroscopic X-Ray System


Hazard / Issue

It may occur that after system startup no stand movement is possible any longer in the event of a discharged BIOS battery of the robotic stand control PC. If this problem occurs during startup, all stand movements are blocked and can only be reactivated by a field service engineer.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: UDI-DI: Artis zeego 04056869010083 Artis Q.zeego 04056869010007
View official government recall