Siemens Medical Solutions USA, Inc recalls Artis zeego (Model no. 10280959) and Artis Q.zeego (Model n
Recalled Product
Artis zeego (Model no. 10280959) and Artis Q.zeego (Model no. 10848283), Interventional Fluoroscopic X-Ray System
Hazard / Issue
It may occur that after system startup no stand movement is possible any longer in the event of a discharged BIOS battery of the robotic stand control PC. If this problem occurs during startup, all stand movements are blocked and can only be reactivated by a field service engineer.