medical Class II Updated 2021-04-14

Fenwal Inc recalls Amicus Exchange Kit - Product Usage: intended for use in the

Recalled Product

Amicus Exchange Kit - Product Usage: intended for use in the collection of blood components and mononuclear cells.


Hazard / Issue

Some lots of Amicus MNC Apheresis and Amicus Exchange kits have leaking centrifuge packs during a procedure on the Amicus Separator using therapeutic protocols only. The defect is a blood leak at the boot, elbow, or in the channel on the separation chamber of the centrifuge pack.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Product Code X6R2339; UDI: 04086000100151; Batch Numbers FA20F08147 (Exp. 06/07/2022), FA20H12145 (Exp. 08/11/2022), FA20J27157 (Exp. 10/26/2022)
View official government recall