Siemens Medical Solutions USA, Inc recalls COHERENCE Oncologist, Model Nos. 07333680 & 07351898
Recalled Product
COHERENCE Oncologist, Model Nos. 07333680 & 07351898
Hazard / Issue
A potential safety issue arises when an offset calculation is followed by a filter operation; correcting the image alignment after this sequence will result in incorrect offset values which could lead to incorrect repositioning of the patient and dose to wrong location.