Siemens Medical Solutions USA, Inc recalls COHERENCE Oncologist, Model Nos. 07333680 & 07351898

Recalled Product

COHERENCE Oncologist, Model Nos. 07333680 & 07351898


Hazard / Issue

A potential safety issue arises when an offset calculation is followed by a filter operation; correcting the image alignment after this sequence will result in incorrect offset values which could lead to incorrect repositioning of the patient and dose to wrong location.

Issued by

FDA

Affected States: CA, FL, GA, IA, MA, MI, NE, NJ, NV, NY, OH, OK, PA, SD, TN, WV
Lot/Code Info: Software versions 2.0.49, 2.0.50, 2.0.51