medical Class II Updated 2025-03-26

CareFusion 303, Inc. recalls BD Pyxis: Anesthesia Station ES, REF: 327, MedStation ES,

Recalled Product

BD Pyxis: Anesthesia Station ES, REF: 327, MedStation ES, REF: 323, MedStation ES Tower, REF: 352, CII Safe ES, REF: 1116-00, CII Safe ES Desktop PC, REF: 1156-00, CII Safe ES Desktop PC, Wired SCNR, REF: 107-254-01, CII Safe ES Desktop PC, W-less SCNR, REF: 107-255-01, Used with Enterprise Server (ES), REF: 1115-00 with affected Enterprise Server software.


Hazard / Issue

Sync timeout may occur between the server and automated dispensing device, due to a server software error which may occur if syncing a large amount of data that takes excessive time, which may cause the dispensing devices to become disconnected from the ES server; dispensing devices will not receive updated information, which may lead to incorrect therapy, and delay in accessing medication.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: BD Pyxis Enterprise Server software versions: 1.7.4.128, 1.7.4.134, 1.7.4.136, 1.7.4.137, All 1.7.1, All 1.7.2, All 1.7.3. REF/UDI-DI: 327/10885403477836, 323/10885403512667, 352/10885403512674, 1116-00/10885403512605