medical Class II Updated 2018-04-25

Cook Inc. recalls MINC Benchtop Incubator (Mini Incubator) catalog (K-MINC-10

Recalled Product

MINC Benchtop Incubator (Mini Incubator) catalog (K-MINC-1000 US) G46022 is a microprocessor controlled, gassed, humidified, incubator intended for use in cell culture.


Hazard / Issue

The temperature of the incubation lids of the MINC Bench top Incubators may become too high or too low for proper gamete/embryo incubation due to fatigue failure of the flat flex lid heater.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: All lots
View official government recall

Looking for safe alternatives?