Cook Inc. recalls MINC Benchtop Incubator (Mini Incubator) catalog (K-MINC-10
Recalled Product
MINC Benchtop Incubator (Mini Incubator) catalog (K-MINC-1000 US) G46022 is a microprocessor controlled, gassed, humidified, incubator intended for use in cell culture.
Hazard / Issue
The temperature of the incubation lids of the MINC Bench top Incubators may become too high or too low for proper gamete/embryo incubation due to fatigue failure of the flat flex lid heater.