medical Class II Updated 2021-04-14

Becton Dickinson & Co. recalls BD BBL Martin-Lewis Agar 20 Ea - Product Usage: use for the

Recalled Product

BD BBL Martin-Lewis Agar 20 Ea - Product Usage: use for the isolation of pathogenic Neisseria from specimens containing mixed flora of bacteria and fungi. Catalog Number: 221557


Hazard / Issue

Products show heavy growth of C. albicans after 24 hours of incubation due to misformulation (expected reaction is partial inhibition after 18-48 hours incubation) and reduce the identification of Neisseria contributing to a delay in time to definitive diagnosis or a delay in proper treatment

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Lot Number: 0317558, Exp. Date 2021-MAR-11 UDI: (01)30382902215579 (10)0317558(17)031121 (30)0020