medical Class II Updated 2022-07-20

Breas Medical, Inc. recalls Vivo 45LS Ventilator- intended to provide continuous or inte

Recalled Product

Vivo 45LS Ventilator- intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation Model Number: 230000 ¿ If the recalled product is a software, include version number: firmware versions 5.0.5 or earlier


Hazard / Issue

A forced shutdown of one of the processors did not generate a watch dog alarm, would develop into a failure AND the instructions to monitor a ventilator dependent patient are NOT followed, the health consequences could potentially be Permanent impairment or life threatening if medical intervention is not obtained

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Unique Device Identifier (UDI): UDI 07321822300004. Serial number ranges in scope are: D******, K******, M01**** to M040124, where asterisk (*) denotes a digit from 0 to 9.
View official government recall