medical Class II Updated 2026-02-25

Auris Health, Inc recalls MONARCH Platform, Bronchoscopy, Catalog Numbers: MON-000005-

Recalled Product

MONARCH Platform, Bronchoscopy, Catalog Numbers: MON-000005-01 , MON-000005-01R , MON-000006 , MON-000006-RFB, MON-000008 with software


Hazard / Issue

Device for bronchoscopic visualization, patient airway access has software issue: if application restarts after patient-side selection, prior to bronchoscope loading, application will re-initialize to patient-left position; and if right position was previously selected, this may lead to unanticipated robotic arm positioning toward patient-left, may result in contact with patient/endotracheal tube.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Software versions: 2.2.2, 2.2.3, 2.2.4, 2.2.5, and 4.1.1. UDI-DI: 10810068810803, 10810068810988. Serial Numbers: 110024, 120088, 120057, 120046, 120058, 120015, 120160, 120068, 120081, 120056, 110028, 120128, 120017, 120089, 120075, 110016, 120031, 120169, 120113, 120168, 120055, 120014, 120060, 120174, 120195, 120026, 110038, 120124, 120194, 110027, 110035, 120123, 120091, 120007, 110022, 120006, 120152, 120004, 120120, 120041, 120106, 110034, 120126, 120082, 120151, 120051, 120020, 120153, 120193, 120170, 120108, 120037, 120074, 120199, 120130, 120095, 110026, 120096, 120009, 120111, 120022, 120054, 120163, 110045, 120027, 110039, 120090, 120141, 120072, 120155, 110049, 120143, 110047, 120125, 120052, 110052, 120100, 120050, 120098, 120045, 120131, 110020, 110030, 110032, 110021, 120013, 120105, 120136, 120192, 110017, 120029, 120139, 110031, 120137, 120053, 110029, 120028, 120034, 120085, 120019, 120049, 120209, 120144, 120161, 120001, 110033, 120145, 120127, 110046, 110043, 120011, 120070, 120076, 120134, 120117, 110025, 120077, 120149, 120191, 120102, 120002, 120156, 120010, 120171, 120003, 120023, 120146, 120188, 120043, 120071, 120042, 120196, 110051, 120065, 120172, 120116, 120150, 120066, 120166, 120033, 110013, 120079, 120008, 120086, 120032, 120073, 1806P1, 120197, 120080, 120078, 14, 120012, 110041, 120084, 120219, 110036, 120181, 120083, 120067, 120021, 120216, 120217, 120218, 120069, 120204, 120207, 120215, 120237, 120208, 120213, 120164, 120132, 120182
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →