Beckman Coulter Inc. recalls UniCel DxH 900 Coulter Cellular Analysis System, REF number

Recalled Product

UniCel DxH 900 Coulter Cellular Analysis System, REF number C11478 Product Usage: Note: The DxH 900 hematology analyzer is CLIA categorized under document CR180294 and is currently marketed as a member of the instrument family of the DxH 800 cleared under K140911, per FDA s guidance, Guidance for Industry and FDA Staff, Replacement Reagent and Instrument Family Policy . DxH 800 intended use states: The UniCel DxH 800 Analyzer is a quantitative multi-parameter, automated hematology analyzer for in vitro diagnostic use in screening patient populations found in clinical laboratories. The UniCel¿ DxH 800 Analyzer identifies and enumerates the parameters indicated below on the following sample types: Whole Blood (Venous and Capillary) o WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, RDW-SD, PLT, MPV, NE%, NE#, LY%, LY#, MO%, MO#, EO%, EO#, BA%, BA#, NRBC%, NRBC#, RET%, RET#, MRV, IRF Pre-Diluted Whole Blood (Venous and Capillary) o WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, RDW-SD, PLT, MPV Body Fluids (cerebrospinal, serous and synovial) o TNC and RBC


Hazard / Issue

Complaints received of sporadic erroneously elevated platelet results without flags or system messages. Thrombocytopenia may go unrecognized.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: All current software versions are impacted