medical Class II Updated 2021-04-21

Covidien, LP recalls Covidien Signia Grey Small Diameter Curved Tip Intelligent R

Recalled Product

Covidien Signia Grey Small Diameter Curved Tip Intelligent Reload 30 mm Vascular 8 mm - Short Item Code: SIGSDS30CTV


Hazard / Issue

Potential for the device safety interlock to fail. The safety interlock prevents an empty single-use reload from being erroneously fired a second time. If the safety interlock fails and the empty single-use reload is fired a second time, the stapler may cut tissue but no staples will deploy from the empty reload, potentially resulting in a failed anastomosis, bleeding, tissue trauma, and the potential need to convert to an open procedure.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Lot Numbers: N0A0191UY N0D0392UY N0F0406UY N0L0708UY N0B0657UY N0E0133UY N0G0907UY N9K0376UY N0C0349UY N0E0956UY N0J0746UY N9K1093UY N0C0492UY N0E1026UY N0K0248UY N9L0796UY N0C1008UY N0F0202UY N0K0803UY N9M0418UY N0D0283UY N0F0221UY N0L0307UY N9M0651UY UDI: 10884521741881 20884521741888
View official government recall

Looking for safe alternatives?